Validation of Respiratory Inductance Plethysmography for Measuring Tidal Volume in Swine  期刊论文  

  • 编号:
    12736aee-911d-4264-bcc5-7088d7f9c805
  • 作者:
    Su, Zhenbo(苏振波)#[1,2,3]Oto, Jun[2,3];Wang, Jingwen(王静雯)[2,3,4]Kimball, William R.[2];Chenelle, Christopher T.[3];Kacmarek, Robert M.[2,3];King, David R.[5];Jiang, Yandong[2];Duggans, Michael J.*[5]
  • 语种:
    英文
  • 期刊:
    COMPARATIVE MEDICINE ISSN:1532-0820 2015 年 65 卷 3 期 (225 - 231) ; JUN
  • 收录:
  • 摘要:

    Measuring tidal volume (V-T) in nonintubated swine or swine with leaking breathing circuits is challenging. The aim of this study was to validate respiratory inductance plethysmography (RIP) for measuring V-T in swine that are comparable in size to adult humans. To determine calibration curves, V-T and RIP readings were obtained from anesthetized swine (n = 8; weight, 46-50 kg) during positive-pressure (mechanical) ventilation and spontaneous breathing. For positive-pressure ventilation, 6 pigs were mechanically ventilated by using the pressure-control mode. The 2 pigs in the spontaneously breathing cohort each received a single intravenous bolus dose of propofol to abolish spontaneous breathing; V-T was measured during gradual return of their respiratory drive. A flow volume sensor was placed between the proximal end of the endotracheal tube and breathing circuit for the recording of inspiratory and expiratory V-T. RIP readings were recorded by using 2 bands, which simultaneously measured ribcage and abdominal excursions. The data revealed that V-T was linearly correlated with the movements of both ribcage and abdomen as measured by using plethysmography over a large range of tidal volume (44 to 1065 mL). In addition, the intercept of the linear equation was small or even negative during spontaneous breathing but increased significantly (maximum, 145 mL, 59.2 +/- 35.1 mL) during positive pressure ventilation. Our results indicate that V-T in swine can be calculated by using a simple univariate linear regression equation with RIP readings obtained during either mechanical ventilation or spontaneous breathing.

  • 推荐引用方式
    GB/T 7714:
    Su Zhenbo,Oto Jun,Wang Jingwen, et al. Validation of Respiratory Inductance Plethysmography for Measuring Tidal Volume in Swine [J].COMPARATIVE MEDICINE,2015,65(3):225-231.
  • APA:
    Su Zhenbo,Oto Jun,Wang Jingwen,Kimball William R.,&Duggans Michael J..(2015).Validation of Respiratory Inductance Plethysmography for Measuring Tidal Volume in Swine .COMPARATIVE MEDICINE,65(3):225-231.
  • MLA:
    Su Zhenbo, et al. "Validation of Respiratory Inductance Plethysmography for Measuring Tidal Volume in Swine" .COMPARATIVE MEDICINE 65,3(2015):225-231.
  • 条目包含文件:
    文件类型:PDF,文件大小:
    正在加载全文
浏览次数:17 下载次数:0
浏览次数:17
下载次数:0
打印次数:0
浏览器支持: Google Chrome   火狐   360浏览器极速模式(8.0+极速模式) 
返回顶部