Realization of fully automated quantification of left ventricular volumes and systolic function using transthoracic 3D echocardiography  期刊论文  

  • 编号:
    6cf94ce6-a9c1-4998-ad5a-f899a6f96b42
  • 作者:
  • 语种:
    英文
  • 期刊:
    CARDIOVASCULAR ULTRASOUND ISSN:1476-7120 2018 年 16 卷 ; JAN 23
  • 收录:
  • 关键词:
  • 摘要:

    Background: Study on automated three-dimensional (3D) quantification of left heart parameters by using Heartmodel software is still in the early stage and fully automatic analysis was not clearly achieved. The aim of our study was to evaluate the performance of this new technology in measuring left ventricular (LV) volume and ejection fraction (EF) in patients with a variety of heart diseases on the basis of rationally determining the default endocardial border values.
    Methods: Subjects with a variety of heart diseases were included prospectively. High quality Heartmodel images were selected to determine the end-diastolic and end-systolic default values of endocardial border. The accuracy and reproducibility of automated three-dimensional echocardiography (3DE) for measuring LV end-diastolic volume (EDV), end-systolic volume (ESV) and EF were evaluated with the traditional manual 3DE as the relative standard.
    Results: Ninety seven subjects were enrolled in the study. The default endocardial border values were determined as 66% and 40% for end-diastole (ED) and end-systole (ES), respectively. Most of the subjects (84/ 97) were automatically analyzed by Heartmodel software without manual adjustment, revealing a close correlation of automated 3DE with manual 3DE in measuring EDV, ESV and EF (r-values: EDV: 0.96, ESV: 0.97, EF: 0.96). The EDV and ESV values obtained by automated 3DE were higher than those measured by manual 3DE (biases: EDV: 16 +/- 18 ml, ESV: 11 +/- 12 ml). The intra-and inter-observer reproducibility of automated 3DE was better than that of manual 3DE. Automated 3DE with manual adjustment showed good consistency with manual 3DE in assessing the impairment degree of systolic function in patients with wall motion abnormalities (n = 58), (Kappa = 0.74, P = 0.00).
    Conclusion: Fully automated 3DE quantification of LV volume and EF could be achieved in most patients. Since automated 3DE was accurate and more reproducible, it could replace the existing manual 3DE technology and be routinely used in clinical practice.

  • 推荐引用方式
    GB/T 7714:
    Sun Lina,Feng Haiyan,Ni Lujia, et al. Realization of fully automated quantification of left ventricular volumes and systolic function using transthoracic 3D echocardiography [J].CARDIOVASCULAR ULTRASOUND,2018,16.
  • APA:
    Sun Lina,Feng Haiyan,Ni Lujia,Wang Hui,&Gao Dongmei.(2018).Realization of fully automated quantification of left ventricular volumes and systolic function using transthoracic 3D echocardiography .CARDIOVASCULAR ULTRASOUND,16.
  • MLA:
    Sun Lina, et al. "Realization of fully automated quantification of left ventricular volumes and systolic function using transthoracic 3D echocardiography" .CARDIOVASCULAR ULTRASOUND 16(2018).
  • 入库时间:
    12/4/2019 7:16:10 PM
  • 更新时间:
    12/4/2019 7:16:10 PM
  • 条目包含文件:
    文件类型:PDF,文件大小:
    正在加载全文
浏览次数:52 下载次数:0
浏览次数:52
下载次数:0
打印次数:0
浏览器支持: Google Chrome   火狐   360浏览器极速模式(8.0+极速模式) 
返回顶部